In this article we assume that both detectors are SPADs of the same kind (both Excelitas SPADs, or both PDM-SPADs, or both tau-SPADs).
If you are using either a proper (see below) 50/50 beamsplitter to split the signal between two detectors or a 100% mirror to direct the signal to one detector and then the other, but still observe very different count rates on the two detectors, this discrepancy could be due to the following reasons:
1. Misalignment of the Lens in Front of the SPAD:
The lens alignment in front of the SPAD may be incorrect. Typically, the detector mounted in reflection requires adjustment because beamsplitters might not all have exactly the same angle. Solution: Follow the “Everyday Alignment” Procedure from the MicroTime 200 manual to correctly align the lens. It might be that also “Fundamental Alignment” is necessary.
2. Beamsplitter Variability:
PicoQuant offers different types of beamsplitters:
Solution: To determine whether the misalignment or the splitter itself is causing the discrepancies, you can perform the following test:
Test Procedure:
If you have followed these instructions, and the light is not split 50/50, then this is likely a property of the beamsplitter. Make sure that is it the correct kind of beamsplitter (see above), and/or contact us for advice.