User Tools

Site Tools


howto:antibunching_measurements

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revisionBoth sides next revision
howto:antibunching_measurements [2023/09/07 11:34] – [Antibunching using SymPhoTime] peterhowto:antibunching_measurements [2023/09/07 11:43] – [Antibunching using SymPhoTime] peter
Line 16: Line 16:
     *with [[Products:PicoHarp 300]] in [[glossary:T3-mode]]     *with [[Products:PicoHarp 300]] in [[glossary:T3-mode]]
         *not possible with SymPhoTime (32-bit version), because in SPT32 the Sync Divider is always active and set to 8         *not possible with SymPhoTime (32-bit version), because in SPT32 the Sync Divider is always active and set to 8
-        *instead, data acquisition is possible with native [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] softwares as well as with [[software:SymPhoTime64]], where the sync divider can be disabled. With a HydraHarp400, don't use use SYNC channel as input for one detector. Use e.g. input channels 1 and 2. SYNC signal must be disconnected. Unfortunately, analysis of HH T3-antibunching data is not implemented in SPT. +        
 +Instead, data acquisition is possible with native [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] softwares as well as with [[software:SymPhoTime64]], where the sync divider can be disabled. 
 + 
 +Using a PicoHarp300, connect one detector to Ch.0 (SYNC) and the other one to Ch.1 
 + 
 +With a HydraHarp400, don't use use SYNC channel as input for one detector. Use e.g. input channels 1 and 2. The SYNC should stay connected. 
 + 
 +Unfortunately, analysis of HydraHarp400 T3-mode antibunching data is not implemented in SPT. 
     *with [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] in [[glossary:T2-mode]]     *with [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] in [[glossary:T2-mode]]
         *measurements are possible with the [[software:SymPhoTime64]]         *measurements are possible with the [[software:SymPhoTime64]]
howto/antibunching_measurements.txt · Last modified: 2023/09/07 11:54 by peter