**Note:** Calibration is a necessary task for FCS measurements. In a calibration procedure, the confocal volume VEff as well as the structural parameter κ (the ratio of z to xy extension) is determined using a reference dye. This is the prerequisite to determine correct diffusion coefficients and concentrations for the sample of interest.\\ | **Note:** Calibration is a necessary task for FCS measurements. In a calibration procedure, the confocal volume VEff as well as the structural parameter κ (the ratio of z to xy extension) is determined using a reference dye. This is the prerequisite to determine correct diffusion coefficients and concentrations for the sample of interest.\\ | ||
ATTO488 (ATTO-Tec, Germany) is a suited dye to calibrate the confocal microscope at ~470 – 490 nm excitation and detection around 500 – 540 nm.\\ | ATTO488 (ATTO-Tec, Germany) is a suited dye to calibrate the confocal microscope at ~470 – 490 nm excitation and detection around 500 – 540 nm.\\ | ||
- | The diffusion coefficient of ATTO488 carboxilic acid has been measured to be 400 µm²/s at 25°C. Published diffusion coefficients of various fluorophores are summarized in the Technical Note " | + | The diffusion coefficient of ATTO488 carboxilic acid has been measured to be 400 µm²/s at 25°C. Published diffusion coefficients of various fluorophores are summarized in the Technical Note " |
Note that the diffusion coefficients of all dyes are temperature dependent.\\ | Note that the diffusion coefficients of all dyes are temperature dependent.\\ | ||
FCS measurements are usually point measurements acquired with very sensitive detectors ([[glossary: | FCS measurements are usually point measurements acquired with very sensitive detectors ([[glossary: | ||
FCS is a technique applicable over a wide range of concentrations, | FCS is a technique applicable over a wide range of concentrations, | ||
- | |||