This tutorial shows step-by-step, how to draw intensity time traces from a single molecule. The script requires a point measurement of an immobilized fluorophore. When applied to images, it can be used to check for line and frame marker presence and assignment.
Note: The “Samples” workspace is delivered with the SymPhoTime 64 and on the CD-ROM and contains example data to show the function of the SymPhoTime data analysis. If you haven't installed it on your computer, copy it from the DVD onto a local drive before going through this tutorial, e.g. to a folder C:\SPT_User
.
Response: The files of the sample workspace are displayed in the workspace panel on the left side of the main window.
Cy5_immo_Lifetime_Trace.ptu
by a single mouse click.Note: This file contains a fluorescence time trace of an immobilized Cy5 molecule, its fluorescence was split by polarization onto two detectors.
Note: The drop down menu can be opened and closed by clicking on the grey button on the left side of the header of the drop down menu:
Response: The “Intensity Time Trace” script is applied to the file Cy5_immo_Lifetime_Trace.ptu
. Thereby, a new Window opens:
Note: The window contains two different regions:
Response: As activated under “Select data channels”, two intensity plots are calculated and plotted according to the two selected detector channels.
Note: The plot consists of several parts:
In the small upper panel, the complete time trace is plotted as the sum of all intensity channels.
The main intensity time trace plots show the subsection of the trace marked in green in the upper trace for the two channels.
On the right, photon counting histograms (PCH) are plotted, indicating the frequency of different intensity values.
Response: A window opens asking for a file name to store the data.
Response: A result file TimeTrace.pqres
is generated.
Note: PicoQuant photon counting boards can register up to four external marker signals. These marker signals are usually generated by a scanning device. Usually, in a FLIM&FCS upgrade kit for Laser Scanning Microscopes (LSM), three markers are used: “line start”, “line stop” and “frame marker”.
These markers can be visualized using the “Intensity Time Trace” script to assign them correctly during the configuration of new setups.
GUVs.ptu
and apply the time trace script to that file. When the intensity time trace window opens, click “Calculate” using a binning of 1 ms as set by default.Note: When the intensity time trace of an image is displayed, the scanning results in intensity spikes.
Response: The marker signals are drawn in the time trace. It can already be seen that the blue marker is more sparse. The marker pull down window also shows, now many markers are present.
Note: The image contains three different markers. Marker 1 and 2 are the most frequent markers, i.e. they serve as line start and line stop signals, whereas marker 3 is less frequent and serves as frame marker. In the image below, the frame markers can be clearly distinguished, while the line markers are too closely spaced for discrimination.
Note: It can be clearly seen that the distance between the black marker 1 and the next red marker 2 to the right is bigger than the distance between the red marker 2 and the following black marker 1 on the right. As this was a mono-directionally scanned image, the return from the end of the line to the line start position is done with highest possible speed. In contrast, the scanning of the actual line is performed according to the user settings, which are usually slower, but never faster than the backfly of the scanner. Therefore, in this measurement, marker 1 and 2 are the line start and line stop markers, respectively.