- | [[HowTo (Experimental)]]; | + | |
===== Q.: How can we avoid decay at very long time delay (up to 5us)? ===== | ===== Q.: How can we avoid decay at very long time delay (up to 5us)? ===== | ||
Line 9: | Line 8: | ||
- detectors: perkin elmer SPCMAQR14 | - detectors: perkin elmer SPCMAQR14 | ||
- | - 700 nm SP filter in front of one of the apds | + | - 700 nm SP filter in front of one of the APDs |
- sample: diamond nanocrystals (100 nm), we are measuring the antibunching of the NV centers emission | - sample: diamond nanocrystals (100 nm), we are measuring the antibunching of the NV centers emission | ||
Line 19: | Line 18: | ||
===== A.: The decay can be caused by blinking of the nanodiamonds. ===== | ===== A.: The decay can be caused by blinking of the nanodiamonds. ===== | ||
==== A decay caused by correlation statistics ==== | ==== A decay caused by correlation statistics ==== | ||
+ | This decay is caused by correlation statistics. This happens because you correlate the first arriving photon against a photon at the longer time, which will be an increasingly unlikely event because the earliest stop photons always win and the experiment restarts. | ||
- | On top of an eventual blinking you will have a decay caused by correlation statistics. | + | I attached |
- | This happens because you correlate the first arriving photon against a photon at the longer time, | + | [{{howto: |
- | which will be an increasingly unlikely event because the earliest stop photons always win and the experiment restarts. | ||
- | I attached a simulation. | ||
- | [{{howto: | ||
- | ==== You will I need the raw data ==== | + | \\ |
+ | ==== You will I need the raw data ==== | ||
- | The timescale is in ns. The plateau is the dead-time. The decay slope depends on the count rate. | + | The timescale is in ns. The plateau is the [[glossary:dead time]]. The decay slope depends on the count rate. |
- | You can zoom into your data and you will see roughly the same behavior - linear decay after the dead-time, which than of course is overshadowed by the background. | + | You can zoom into your data and you will see roughly the same behavior - linear decay after the [[glossary:dead time]], which than of course is overshadowed by the background. |
Line 42: | Line 40: | ||
For a trial you will I need the raw data - a pt2 file containing all the photon data - not just the correlated curve. You can record this using the button I highlighted in the attached screenshot. | For a trial you will I need the raw data - a pt2 file containing all the photon data - not just the correlated curve. You can record this using the button I highlighted in the attached screenshot. | ||
+ |