~~TOC~~
Diamond NV Centers
Basics
Excitation
Emission
Figure 3 shows the emission spectra of a NV-center. The pronounced Peak at 637nm is the zero-phonon line of the diamond lattice (comparable to Raman).
Fluorescence Lifetime
Measuring NV centers
Figure2: Jablonski diagram of the luminescent transition of the nitrogene-vacancy in a diamond lattice. Bright fluorescence corresponds to the ^3E→^3A (m_s=0) transition
Due to their photo-physical stability the NV-ceneters make for excellent demonstration samples for antibunching measurements. They are also potential candidates to image the confocal volume.
The following will explain how to find and measure NV-centers in bulk diamond.
System setup
refractive index of diamond is 2.4. that means oil immersion is necessary as well as a high N.A.
used objective: N.A. 1.3 is working, better 1.45 oil!
Although measurements were possible also with the water immersion objective: with 60×1.2 water less than 50% count rate in comparison to 100×1.3 oil
0.95 N.A. air objective did not yield any usable results.
excitation source: 532 nm
excitation power: >7000 a.u. @ 40MHz, > 2000 a.u. @ 10MHz (basically as high as possible)
major dichroic: 532 / 635
-
objective: 1.3 N.A. oil immersion
Figure3: fluorescence spectrum of a single NV center in diamond excited with 532 nm
Mounting the sample
sample is mounted directly on top of the objective (with immerision oil, without any coverslips etc.).
Imaging/ finding nv centers
Figure4: Prescan of NV-centers in bulk diamond. 5 micrometer into the sample, 80×80 micrometer, 150×150 pixel resolution. 532nm excitation, 690/70 emission, 40MHz rep. rate, approx. 7000 a.u.
move approx. 5 micrometer into the sample.
Although photo-physical stability makes the nv-centers ideal single emitter samples compared to single molecules the laser power to excite them has to be considerably higher. (0.1 mW and above).
Figure 4 shows a prescan of the complete field of view. The orange dots are NV-centers.
zoom into a region with a few NV-centers (see figure 5),
select one, and position the focus on the NV-center.
use the oscilloscope and pifoc+piezo to move the NV-center in x,y and z into the center of the focus.
For lifetime analysis switch laser repetition rate to 10MHz (see figure 6)
check whether the fluorescence lifetime is around 12 ns.
the NV-center's emission should stay stable (no bleaching)
Figure5: NV-centers in bulk diamond. 5 micrometer into the sample, 532nm excitation, 690/70 emission, 40MHz rep. rate, 6789.96a.u., 2ms/pixel, 15×15 micron
Typical Measurement Results
when measuring the fluorescence intensity of a single NV center, the count rate should stay stable. The time trace in figure 8 shows a reduction of the countrate due to stage drift (temperature etc, or in this case probably because the MT200 inhouse II was mounted on a wheelbarrow instead of an optical table)
to perform
antibunching measurements on a selected NV-center choose a NV-center that is isolated, measuring time of 120 s should be enough to yield an antibunching measurement comparable to figure 9.
NOTE: Measurement setup has to be changed (see antibunching measurements)
Figure6: same as figure 5 but recorded with 10
MHz rep rate., 2304.94 a.u.;4 ms/pixel 45 min rec. time!
Figure7: Fluorescence Lifetime Fit of a selected NV center, 10 Mhz reprate, Inhouse II, MPD/SPAD
Figure8: No bleaching visible. reduction in count rate is due to mechanical drift (inhouse II) on wheelbarrow
Figure9: Antibunching measured on NV-center. Acquisition time 120 s, 1.3 N.A. oil immersion, 3534.98 a.u. 532 nm 690/70 emmission filters
References
M. Boersch, R. Reutera, G. Balasubramaniana, R. Erdmann, F. Jelezkoa, J. Wrachtrup Fluorescent nanodiamonds for FRET-based monitoring of a single biological nanomotor FoF1-ATP synthase Proc. of SPIE Vol. 7183, 71832N (2009)
http://dx.doi.org/10.1117/12.812720